Anti- Parkinsonian Drug Estimation by RP-HPLC

Abhinandana, Patchala and Nadendla, Rama Rao (2021) Anti- Parkinsonian Drug Estimation by RP-HPLC. Journal of Pharmaceutical Research International, 33 (1). pp. 14-25. ISSN 2456-9119

[thumbnail of 1980-Article Text-3609-3-10-20221006.pdf] Text
1980-Article Text-3609-3-10-20221006.pdf - Published Version

Download (406kB)

Abstract

Aim: The main aim of the current study is to give best and simple method for the estimation of antiparkinsonian drugs named Carbidopa, levodopa and entacapone.

Study Design: Simultaneous estimation of Carbidopa, levodopa and entacapone was performed by using Quadrapumped (SHIMADZU Prominance-i, LC-2030C) RP-HPLC equipped with PDA detector.

Place and Duration of Study: Chalapathi Drug Testing Laboratory, Chalapathi Institute Of Pharmaceutical Sciences, Lam, Guntur-522034, Andhra Pradesh, India during the period of August 2019 to February 2020.

Methodology: The assets of the study can determined as the process of qualification and quantification was done on SHIMADZU Prominance-i, LC-2030C system equipped with Phenomenex ODS (150 x 4.6 mm, 5μm) column and mobile phase was optimized using combination of acetonitrile and 0.1% ortho phosphoric acid in the ration of 50:50 v/v at a flow rate 1.0 ml/min. The wavelength was set as 270nm at ambient temperature by injecting 20µl of solution and the run time was fixed for 5 min.

Results: Calibration plot shown best regression over the concentration range of 5-160 µg/ml of Carbidopa, Levodopa and Entacapone standard solutions. The LOD and LOQ were found to be 0.85 and 2.54 µg/ml for Entacapone, 0.24 and 0.71 µg/ml for Levodopa, 0.14 and 0.43 µg/ml for Carbidopa respectively. The accuracy of the proposed method was determined by performing recovery studies and was found to be between 98-102%. The repeatability testing for both sample and standard solutions was found as %RSD<2.0% which is within the acceptable limits showing that the method is precise as well. The proposed method was successfully applied for the marketed formulations of Carbidopa, Levodopa and Entacapone tablets. In addition the main feature of proposed method is economic and eco-friendly with less retention time around 5.0 min.

Conclusion: Including all the optimized method parameters and statistical results given it can be concluded as a new, simple, sensitive, precise and accurate economical analytical method was developed and validated by RP-HPLC for the detection and quantification of Carbidopa, Levodopa and Entacapone which can be applied to the marketed formulation where there are no official compendial methods reported for this particular combination. The high sensitivity (LOD), mobile phase utilized and run time (=5) can be determined as an important features for this proposal.