Automated phosphor thermometry lifetime calibration of multiple phosphors and emission lines to above 1900 K

Feuk, Henrik and Nilsson, Sebastian and Richter, Mattias (2022) Automated phosphor thermometry lifetime calibration of multiple phosphors and emission lines to above 1900 K. Measurement Science and Technology, 33 (12). p. 127003. ISSN 0957-0233

[thumbnail of Feuk_2022_Meas._Sci._Technol._33_127003.pdf] Text
Feuk_2022_Meas._Sci._Technol._33_127003.pdf - Published Version

Download (1MB)

Abstract

A method is specified which enables lifetime calibration of multiple phosphors and emission lines at the same time to temperatures above 1900 K. The experimental setup and algorithm used for data collection and experimental equipment control are described. The phosphors were coated on an alumina oxide disc and the reference temperature was measured using three type B thermocouples. The algorithm automates the data collection process such that no input from an operator is required during operation. The potential systematic error in calibration temperature was evaluated and was less than 1% around 1400 K.